Are there any examples about how to identify the accuracy of LED measuring instrument?

Menu

Best Sale

Are there any examples about how to identify the accuracy of LED measuring instrument?

Someone provides us with a case for a portable LED verification instrument with touch panel (item#XX-350). We’d like to share you this case and make a simple comparison of the accuracy of the results are as follows: the accurate peak wavelength (lambda P) of general fluorescent lamp is located in 435.833nm, but if the light intensity of LED measuring instrument do not be calibrated, then the peak wavelength (lambda P) is located in 546.074nm. (If the correction of light intensity didn’t be done, color rendering (CRI), temperature (CCT), color coordinates, luminous flux... Wait for measured values… can not be calculated.)

Please refer to the right illustration about the testing results measured by the Rainbow-Light portable LED measuring instrument: checking peak wavelength (lambda P) value of 435.8nm, 435.833nm, 546.074nm and 404.656nm, you can find the measured values of 404.852nm, 435.805nm, 545.673nm are very accurate.

The portable LED measuring instrument XX-350 with the touch panel version is measured below:
The peak wavelength (lambda P) value of 544nm (even if it is 545.673nm, spectral deviation is far more than 1nm), 404.656nm, 435.833nm measured, the spectral deviations are more than 6nm. Thus, the portable LED measuring instrument XX-350 is unable to measure color rendering (CRI). Color temperature (CCT), color coordinates, luminous flux....accurately.

related products
  • Other Customized Specification - Rainbow-Light All-rounder Spectrometer  and other customized Specification, such as cooling、near-infrard、Wifi Spectrometers.
    Other Customized Specification
    Rainbow Series, Aurora Series, Volcano Series, Lightning Series

    Rainbow-Light introduced Super Spectrometer and its customerized specification: the wavelength range of 180 ~ 1700 nm, UV-VIS-NIR and provide a suitable resolution according to customers' demands. Exclusive technology of Dynamic Thermal Equilibrium (DTE) as to has the features of thermal stability, anti-vibration and resolve the unstability of traditional spectrometer with complicated optical components. We provide various customized resolutions of Micro-Spectral Measurement and application for the wavelength range of 180 ~ 1700 nm.